: Portland Personal Injury LawyerCertain Models of Triple-Channel Infusion Pumps Recalled
By Jenny Albano
On Friday, the FDA alerted healthcare professionals of a class 1 recall of certain models of an upgraded triple-channel volumetric infusion pump made by Colleague and Colleague CX. A software malfunction may sound an alarm in the device, showing an error code 16:310:867:0002, which will stop infusing.
An electronic infusion pump delivers controlled amounts of medicine or other fluids to patients through a direct line into the bloodstream (intravenously, intra-arterially, or through an epidural).
"The error occurs when all three channels are in use and a healthcare professional initiates a user interface-intensive command such as overriding the Guardian dose limit feature, displaying the dose mode window, performing dose calculations or performing running rate changes," the company says in a news release.
So far there have been sixteen injuries reported. All the affected users have been contacted and have had their infusion pumps replaced with alternate devices, such as non-upgraded triple-channel and upgraded single-channel pumps.
From May 14 to June 18, 2007 there have been 4512 defective infusion pumps administered to 427 customers, according to an alert from the FDA's MedWatch.
For more information on this subject, please refer to our section on Drugs, Medical Devices, and Implants.
Full post as published by Portland Personal Injury Lawyer on July 25, 2007 (boomark / email).
Baxter Electronic Infusion Pumps
FDA Issues Urgent Class I Recall
E-Z-G0 Golf Cars
RXV Models Recalled due to Injuries
Triple Play Products
Recalls Sit 'N' Stroll Child Restraint Systems
BMW Faces Class Action over High Pressure Fuel Pumps
BMW Faces Class Action over High Pressure Fuel Pumps
Medtronic
Recalls Specific Lots of Paradigm Quick-Set Infusion Sets
Donjoy Inc. and DJ Ortho, Canada Inc.
alleging pain pumps are linked to chondrolysis.

Medtronic SynchroMed EL Implantable Infusion Pumps Recalled
Class 1 Recall of Baxter COLLEAGUE and FLO-GARD Volumetric Infusion Pumps
Baxter Expands Recall of Infusion Pump Class 1 Recall
Medtronic SynchroMed EL Implantable Infusion Pump
FDA Issues Recall of Medtronic Infusion Pump